CellTank-34 port layout

CellTank is available with 900 mm long weldable silicone hose connection.

Con

Purpose

Feature

1

Access with swabable valve

Bottom sample port

2

Access between pump and matrix

Inlet, inoculation

3

Access after matrix, valve, 2 x swabable valve

Product, media out

4

Liquid level control

level 1,300 ml height

5

Head space liquid inlet

base

6

Head space gas inlet, 50 mm sterile filter

<50 l/h,<50 mBar

7

Sparging gas inlet, 50 mm sterile filter

<50 l/h,<50 mBar

8

Integrated non-invasive well for thermo couple, max OD 6 mm sensor

9

Exhaust via twin 50 mm sterile filter or a 500 ml foam collection vessel follow by one OD50 mm sterile filter.

10

PG13.5 extra port, closed with threaded plug

11

PG13.5 port closed with threaded plug, for pH sensor, may also be supplied pre-installed, pre-sterilized

12

PG13.5 port closed with threaded plug, for dO2 sensor, may also be supplied pre-installed, pre-sterilized

13

PG13.5 port for SUS bio mass sensor supplied pre-installed

14

Short tube with swabable valve

Matrix sample

15

Dual liquid level sensor, solid insulated SS rods at 1,300 ml height

16

System back pressure, WC connection via hose to OD 25 mm sterile filter

Photos illustrate the stainless steel support rod, which support

  1. WC hose and blue sterile filter for pressure measurements from port 16
  2. One exhaust filters from port 9
  3. One inlet sterile filter for sparging gas port 7
  4. One inlet sterile filter for head space flushing on port 6

Photo illustrate 500 ml foam collection bottle next to SUB.